Home |
Search |
Today's Posts |
![]() |
|
Electronics Repair (sci.electronics.repair) Discussion of repairing electronic equipment. Topics include requests for assistance, where to obtain servicing information and parts, techniques for diagnosis and repair, and annecdotes about success, failures and problems. |
Reply |
|
|
LinkBack | Thread Tools | Display Modes |
#1
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
I have a Sony CDP-302 player
this is an old, ~1986, high end player. It playes very well --- as long as it is tipped to and angle of 45° or greater on either side. Not 40°, but 45°, maybe 43°. I measured it. If I tip it from 40° to 45° it will start working. No bumping or jarring required. If it is flat or tipped to any angle front to back, the spindle motor won't operate. The spindle motor circuit doesn't get any drive signal. I traced it back to the electronics board and something is shutting down the generation of the signal. The problem is on the CD carriage, because I loosened the electronics board and tipped them seperately. The spindle motor is an open DC type. Coils are on the PC board with the magnet attached to the bottom of the spindle. NOTHING appears to be loose or jammed. All micro switches seem to be operating in all positions. It worked intermittently in the flat position for a while. It is suppose to spin the CD on insertion or power on and read the TOC, then stop and wait for the play button. It does this only when tipped at the proper angle. I don't have a schematic, but what would be position sensitive to shut down the drive signal? thanks |
#2
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() wrote in message ... I have a Sony CDP-302 player this is an old, ~1986, high end player. It playes very well --- as long as it is tipped to and angle of 45° or greater on either side. Not 40°, but 45°, maybe 43°. I measured it. If I tip it from 40° to 45° it will start working. No bumping or jarring required. If it is flat or tipped to any angle front to back, the spindle motor won't operate. The spindle motor circuit doesn't get any drive signal. I traced it back to the electronics board and something is shutting down the generation of the signal. The problem is on the CD carriage, because I loosened the electronics board and tipped them seperately. The spindle motor is an open DC type. Coils are on the PC board with the magnet attached to the bottom of the spindle. NOTHING appears to be loose or jammed. All micro switches seem to be operating in all positions. It worked intermittently in the flat position for a while. It is suppose to spin the CD on insertion or power on and read the TOC, then stop and wait for the play button. It does this only when tipped at the proper angle. I don't have a schematic, but what would be position sensitive to shut down the drive signal? thanks That type of motor normally has many connections to the stator coils, which are a comparitively heavy assembly in themselves. I would suggest that the first move would be to resolder all of those connections, if you can even get at them. Arfa |
#3
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
I agree with Arfa, sounds like loose solder somewhere on the wires or
on the pcb. Sony also had problems with ribbon cables. Sparkey |
#4
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
Its more likely the laser focussing circuit is off somehow, and only
actually locks onto the disc with the mechanical lean to one side. A gentle!! adjustment of laser power, Fgain and Foffset (remembering their original places) may help. If however there are squeeking noises from the laser while the disc doesn't spin then its possible you have a spindle drive problem. "Sparkey" wrote in message ups.com... I agree with Arfa, sounds like loose solder somewhere on the wires or on the pcb. Sony also had problems with ribbon cables. Sparkey |
#5
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() "Cliff Top" wrote in message ... Its more likely the laser focussing circuit is off somehow, and only actually locks onto the disc with the mechanical lean to one side. A gentle!! adjustment of laser power, Fgain and Foffset (remembering their original places) may help. If however there are squeeking noises from the laser while the disc doesn't spin then its possible you have a spindle drive problem. Sorry Cliff, but having worked on Sony CD players since they first came out, I would absolutely dispute that your scenario is " more likely ". Focus servos will lock a normal focus motor ( and indeed likewise the tracking servo will do the same for the tracking motor ) at ANY angle, including upside down. The lens, lens suspension and tracking and focus motors, do not have enough mass to be sufficiently affected by gravity, that any such effect is not easily compensated for by the power of these motors. I can't remember *ever* having had a situation where focus gain and offset adjustments have drifted on their own to any degree anyway, and still less, have caused a problem that could be corrected by imposing a mechanical tilt to a deck. Also, I would absolutely advise against adjusting the laser power - not even gently. It is NOT a valid adjustment on any Sony player, and in fact, on very few players of any other manufacture, Pioneer being a notable exception. Indiscriminate adjustment of laser power - even by a *tiny* amount, can readily result in destruction of an otherwise good laser. Focus gain and focus offset should only be adjusted with the benefit of a service manual, and the proper equipment, or by long experience. The comment about being able to hear the servos squealing, if you rock or rotate the disc by hand when it's not turning under it's own steam however, is a good pointer. If you do hear the squeal, there's a good chance that the focus and tracking servos are locked, and the spindle motor is just not being driven. If there is no servo squeal, then probably, the servos are not locked, which is causing the system control to inhibit the spindle servo. I would still be looking at the spindle motor stator connection soldering first, and also at any connectors and wires going to the spindle motor board. Arfa |
#6
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
Hi!
Also, I would absolutely advise against adjusting the laser power - not even gently. It is NOT a valid adjustment on any Sony player, and in fact, on very few players of any other manufacture, Pioneer being a notable exception. Okay...not that I disagree or anything like that, but why is this adjustment provided if you're not supposed to use it when servicing the player? Why spend the money to include such a thing? Is it for "factory setup only"? William |
#7
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() "William R. Walsh" m wrote in message news:Ul0mh.334766$1i1.84177@attbi_s72... Okay...not that I disagree or anything like that, but why is this adjustment provided if you're not supposed to use it when servicing the player? Why spend the money to include such a thing? Is it for "factory setup only"? Lots of adjustments are critical in various electronics. What about high voltage regulation adustments in CRT based sets. Would you just randomly tweak it up if the pix was weak? Most LP adjustments are not going to do much beyond shortening the life of the pickup. Leonard -------------------------------------------------------------------------------- I am using the free version of SPAMfighter for private users. It has removed 19081 spam emails to date. Paying users do not have this message in their emails. Try SPAMfighter for free now! |
#8
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
Hi!
Lots of adjustments are critical in various electronics. You'd get no argument from me. I know the critical nature of adjustments in electronic devices, and even understand why some of them are not made to be adjusted. But that's not really answering what I asked. I suppose it is possible that I'm asking a question that only CD player manufacturers could answer. :-) I'm not wanting to come across as impolite. I'd simply like to know why this is done. William |
#9
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() "William R. Walsh" m wrote in message news:jU0mh.334794$1i1.210997@attbi_s72... Hi! Lots of adjustments are critical in various electronics. You'd get no argument from me. I know the critical nature of adjustments in electronic devices, and even understand why some of them are not made to be adjusted. But that's not really answering what I asked. I suppose it is possible that I'm asking a question that only CD player manufacturers could answer. :-) I'm not wanting to come across as impolite. I'd simply like to know why this is done. William Your question is not impolite, and I'm sure would not be taken as such by anyone on here. Asking is how you find out. The manufacturing tolerances of the optical block itself, are subject to an amount of spread. These are compensated for when the block is manufactured, by putting it into a test jig that sets certain conditions that should result in a certain range of drive currents to produce a specific level of laser diode output. The pot which sets the drive current is adjusted to achieve what the manufacturers consider to be ideal for that particular laser diode in that particular model of optical block. Once this setting has been made, the pot is very firmly sealed, not just to make it vibration proof, but to indicate that it should not be adjusted from its set position. This is the case for almost all optical blocks. Other characteristics that may result from mechanical tolerances of the deck that the optical block becomes fitted to, or tolerances in the circuitry that the block has to work with, are often provided with adjustments which are intended to be moved as part of a servicing procedure. Examples of these are focus offset, where the zero point of the servo is set to match the average mechanical zero focus point, which should of course be the mechanical resting point of the lens suspension, but may not be if the turntable height is not set exactly correctly. Likewise, tracking offset does the same to compensate for mechanical errors in the tangential or tilt directions. There may also be adjustments which set the gains of the servos to take into account spreads in the servo circuitry. However, many players from the last 10 years or so, have no service physical adjustments at all. With these players the chipsets used have fully electronic internal adjustment systems. As I said in the previous post, one notable exception to the "don't touch it !" rule, is Pioneer players. The laser power pot is a minuscule little thing located on the laser flexiprint, and is adjusted as part of the setup procedure. Pioneer provide a set of testpoints in a single row on their players, and detailed instructions in their manuals, on how to adjust the entire pot-set, which is often a total of eight pots, and which includes the power pot. Interestingly, Pioneer lasers very seldom fail or wear out. I think in 30 odd years, I have only ever replaced perhaps two of the original types ( there were several, but all pretty similar looking ) and maybe two or three of the later ones which have a couple of tiny pots on their PCB. Arfa |
#10
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() "Arfa Daily" wrote in message ... As I said in the previous post, one notable exception to the "don't touch it !" rule, is Pioneer players. The laser power pot is a minuscule little thing located on the laser flexiprint, and is adjusted as part of the setup procedure. Pioneer provide a set of testpoints in a single row on their players, and detailed instructions in their manuals, on how to adjust the entire pot-set, which is often a total of eight pots, and which includes the power pot. Interestingly, Pioneer lasers very seldom fail or wear out. I think in 30 odd years, I have only ever replaced perhaps two of the original types ( there were several, but all pretty similar looking ) and maybe two or three of the later ones which have a couple of tiny pots on their PCB. Arfa This is an interesting point about the Pioneer pickups. Ican't recall ever changing more than one or two myself. I worked at a store that sold tons of Pioneer (Elite, mostly) and since then have not workde on them very much. I can recall cleaning lots of spindle motors, but rarely, if ever changing pickups, Is this the experience of others? What would you attribute the difference in reliability to? Pioneer has always had very good laser products in terms of performance. Do others percieve the reliability to have been very good relative to their competitors as I do? Are the current and lower end products holding up as well? Leonard -------------------------------------------------------------------------------- I am using the free version of SPAMfighter for private users. It has removed 19084 spam emails to date. Paying users do not have this message in their emails. Try SPAMfighter for free now! |
#11
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
I have run thru the solders on the drive board and elsewhere.
The drive is fine when it runs and the "tilt" is a specific angle and repeatable with bumping and jarring having no effect. I am suspecting that it has something to do with the laser itself since the fix (tilt) is so repeatable. It is definitely on the deck. I checked out all the switches and the board on the deck removing and replacing all connectors. Apparently the electronics need feedback from the laser to tell it there is a CD in the deck before it will generate the drive for the spindle. Is this correct??? Therefore the Laser needs to see a reflection. So if the laser is tipped, the reflection won't hit the right spot on the assy. I see only a movement, apparently controlled by coils, that adjusted the laser along the radius of the CD. When it does play, I see the sled shaking back and forth as if following the CD. Shouldn't the laser itself be moving by the coils mentioned above. I think this deck has three lasers to control tracking. I have no idea how this works. I only see one lens. thanks for the responses. On Sun, 31 Dec 2006 10:43:44 GMT, "Arfa Daily" wrote: wrote in message .. . I have a Sony CDP-302 player this is an old, ~1986, high end player. It playes very well --- as long as it is tipped to and angle of 45° or greater on either side. Not 40°, but 45°, maybe 43°. I measured it. If I tip it from 40° to 45° it will start working. No bumping or jarring required. If it is flat or tipped to any angle front to back, the spindle motor won't operate. The spindle motor circuit doesn't get any drive signal. I traced it back to the electronics board and something is shutting down the generation of the signal. The problem is on the CD carriage, because I loosened the electronics board and tipped them seperately. The spindle motor is an open DC type. Coils are on the PC board with the magnet attached to the bottom of the spindle. NOTHING appears to be loose or jammed. All micro switches seem to be operating in all positions. It worked intermittently in the flat position for a while. It is suppose to spin the CD on insertion or power on and read the TOC, then stop and wait for the play button. It does this only when tipped at the proper angle. I don't have a schematic, but what would be position sensitive to shut down the drive signal? thanks That type of motor normally has many connections to the stator coils, which are a comparitively heavy assembly in themselves. I would suggest that the first move would be to resolder all of those connections, if you can even get at them. Arfa |
#12
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
"Leonard Caillouet" writes:
"William R. Walsh" m wrote in message news:Ul0mh.334766$1i1.84177@attbi_s72... Okay...not that I disagree or anything like that, but why is this adjustment provided if you're not supposed to use it when servicing the player? Why spend the money to include such a thing? Is it for "factory setup only"? Lots of adjustments are critical in various electronics. What about high voltage regulation adustments in CRT based sets. Would you just randomly tweak it up if the pix was weak? Most LP adjustments are not going to do much beyond shortening the life of the pickup. The laser power is set at the factory. Only a few components determine the laser power and they don't really change much with age or use. The feedback circuit maintains the laser power constant should the output from the laser diode itself with respect to current descrease. One thing that can affect it though is that dust, dirt, contamination on the optics between the laser and the disc and photodiode can result in less optical power reaching the disc and/or being reflected from the disc to the photodiode. In this case, if the optics can't be cleaned - which is likely for some of it - turning up the laser power is the only real option. But it's a last resort, not a first resort! As has been stated, that's a good way to kill a perfectly good laser. --- sam | Sci.Electronics.Repair FAQ: http://www.repairfaq.org/ Repair | Main Table of Contents: http://www.repairfaq.org/REPAIR/ +Lasers | Sam's Laser FAQ: http://www.repairfaq.org/sam/lasersam.htm | Mirror Sites: http://www.repairfaq.org/REPAIR/F_mirror.html Important: Anything sent to the email address in the message header above is ignored unless my full name AND either lasers or electronics is included in the subject line. Or, you can contact me via the Feedback Form in the FAQs. |
#14
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
Leonard Caillouet wrote:
"Arfa Daily" wrote in message ... As I said in the previous post, one notable exception to the "don't touch it !" rule, is Pioneer players. The laser power pot is a minuscule little thing located on the laser flexiprint, and is adjusted as part of the setup procedure. Pioneer provide a set of testpoints in a single row on their players, and detailed instructions in their manuals, on how to adjust the entire pot-set, which is often a total of eight pots, and which includes the power pot. Interestingly, Pioneer lasers very seldom fail or wear out. I think in 30 odd years, I have only ever replaced perhaps two of the original types ( there were several, but all pretty similar looking ) and maybe two or three of the later ones which have a couple of tiny pots on their PCB. Arfa This is an interesting point about the Pioneer pickups. Ican't recall ever changing more than one or two myself. I worked at a store that sold tons of Pioneer (Elite, mostly) and since then have not workde on them very much. I can recall cleaning lots of spindle motors, but rarely, if ever changing pickups, Is this the experience of others? What would you attribute the difference in reliability to? Pioneer has always had very good laser products in terms of performance. Do others percieve the reliability to have been very good relative to their competitors as I do? Are the current and lower end products holding up as well? Leonard ----------------------------------------------------------------------- One reason Pioneer pickups were so reliable is that the pickup was mounted "upside down" on many models, thus very little dust etc would settle in on the turning mirror. Still, even their models which played the disc label-up were very reliable so far as pickups were concerned. Except of course for those which had the lens fall off... Mark Z. |
#15
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
SAM
Excellent info! Thanks - This assy appears to have the Laser and photodiode array swapped position wise - IE array is at rt angle to the lens. But that doesn't tell me why I can start the CD in a tilted position then lay it flat and it will play just fine till I stop it and try to restart it. It will only START in the tilted position, but will continue to play in any position. It also must be tilted on power up or insertion to read the TOC, otherwise it won't start on "PLAY" at all. I just can't picture anything inside the laser assy that could be that free to move and also be controlled to lock it into position so that it would continue to play once started. The lens assy only rotates on the plane of the CD and doesnt appear to have movement in the "tip" position. DOES the array need to see anything to cause it to start? DOES it search at all by adjusting the tracking of the lens or the array to FIND the CD? DOES the array adjust ITSELF to provide tracking? If so how? thanks On 01 Jan 2007 09:04:51 -0500, Sam Goldwasser wrote: writes: I have run thru the solders on the drive board and elsewhere. The drive is fine when it runs and the "tilt" is a specific angle and repeatable with bumping and jarring having no effect. I am suspecting that it has something to do with the laser itself since the fix (tilt) is so repeatable. It is definitely on the deck. I checked out all the switches and the board on the deck removing and replacing all connectors. Apparently the electronics need feedback from the laser to tell it there is a CD in the deck before it will generate the drive for the spindle. Is this correct??? Therefore the Laser needs to see a reflection. So if the laser is tipped, the reflection won't hit the right spot on the assy. I see only a movement, apparently controlled by coils, that adjusted the laser along the radius of the CD. When it does play, I see the sled shaking back and forth as if following the CD. Shouldn't the laser itself be moving by the coils mentioned above. I think this deck has three lasers to control tracking. I have no idea how this works. I only see one lens. See the CD Player Repair guide at the site below. The sled should not be shaking back and forth, only the lens. The sled should just move occasionally to maintain the lens centered withint its range. --- sam | Sci.Electronics.Repair FAQ: http://www.repairfaq.org/ Repair | Main Table of Contents: http://www.repairfaq.org/REPAIR/ +Lasers | Sam's Laser FAQ: http://www.repairfaq.org/sam/lasersam.htm | Mirror Sites: http://www.repairfaq.org/REPAIR/F_mirror.html Important: Anything sent to the email address in the message header above is ignored unless my full name AND either lasers or electronics is included in the subject line. Or, you can contact me via the Feedback Form in the FAQs. |
#16
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() "Mark D. Zacharias" wrote in message . net... Leonard Caillouet wrote: "Arfa Daily" wrote in message ... As I said in the previous post, one notable exception to the "don't touch it !" rule, is Pioneer players. The laser power pot is a minuscule little thing located on the laser flexiprint, and is adjusted as part of the setup procedure. Pioneer provide a set of testpoints in a single row on their players, and detailed instructions in their manuals, on how to adjust the entire pot-set, which is often a total of eight pots, and which includes the power pot. Interestingly, Pioneer lasers very seldom fail or wear out. I think in 30 odd years, I have only ever replaced perhaps two of the original types ( there were several, but all pretty similar looking ) and maybe two or three of the later ones which have a couple of tiny pots on their PCB. Arfa This is an interesting point about the Pioneer pickups. Ican't recall ever changing more than one or two myself. I worked at a store that sold tons of Pioneer (Elite, mostly) and since then have not workde on them very much. I can recall cleaning lots of spindle motors, but rarely, if ever changing pickups, Is this the experience of others? What would you attribute the difference in reliability to? Pioneer has always had very good laser products in terms of performance. Do others percieve the reliability to have been very good relative to their competitors as I do? Are the current and lower end products holding up as well? Leonard ----------------------------------------------------------------------- One reason Pioneer pickups were so reliable is that the pickup was mounted "upside down" on many models, thus very little dust etc would settle in on the turning mirror. Still, even their models which played the disc label-up were very reliable so far as pickups were concerned. Except of course for those which had the lens fall off... Mark Z. Hi Mark Have you ever managed to replace a completely missing lens, with one from a different pickup ? I had quite an argument with someone once, who said that it just can't be done, but I do it all the time - I should think at least 4 or 5 a year - did one just 2 weeks ago. I use a lens from a scrap KSS series pickup. I think that the trick to getting it to work - or indeed even replacing the 'real' lens when it's still rolling around in the bottom, is to make sure that //every last trace// of glue is removed from both the lens suspension, and the underside of the lens lip, and then to make certain that the lens is refitted absolutely flat. I then refix it, by putting a couple of tiny drops of cyanoacrylate superglue onto the tip of a tiny flat-end jeweller's screwdy, and letting it seep into the crack by capilliary action. Sometimes, I've found that the focus bias and offset need to be adjusted slightly to optimise the performance back to normal, but mostly, I've found that it just works right off, with no improvement to be gained by adjustment. I'd be interested to know what others' experiences of dealing with this problem are. Arfa |
#17
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() "Sam Goldwasser" wrote in message ... writes: I have run thru the solders on the drive board and elsewhere. The drive is fine when it runs and the "tilt" is a specific angle and repeatable with bumping and jarring having no effect. I am suspecting that it has something to do with the laser itself since the fix (tilt) is so repeatable. It is definitely on the deck. I checked out all the switches and the board on the deck removing and replacing all connectors. Apparently the electronics need feedback from the laser to tell it there is a CD in the deck before it will generate the drive for the spindle. Is this correct??? Therefore the Laser needs to see a reflection. So if the laser is tipped, the reflection won't hit the right spot on the assy. I see only a movement, apparently controlled by coils, that adjusted the laser along the radius of the CD. When it does play, I see the sled shaking back and forth as if following the CD. Shouldn't the laser itself be moving by the coils mentioned above. I think this deck has three lasers to control tracking. I have no idea how this works. I only see one lens. See the CD Player Repair guide at the site below. The sled should not be shaking back and forth, only the lens. The sled should just move occasionally to maintain the lens centered withint its range. I agree with Sam. As to whether the player needs to see a disc on the turntable to spin it up, this depends on the unit. Some players will just go ahead and turn on the laser, and spin up the disc, as soon as a 'laser home' signal is detected. Some coarse speed adjustments then take place, along with focus seek, until the spindle servo starts to see valid data coming off the disc, whereupon, it will lock, and the rotational speed will stabilize to the correct value. Other players will switch on the laser as soon as the home signal is detected, and go ahead and look for a reflection from a disc. Coarse focus search will take place at the same time, until the focus servo locks. At this point, the drive to the spindle motor will be switched on, and the data search will commence. Some players, if they fail to detect an initial reflection, will pulse the spindle motor to rotate the disc some, in case there was a jammy fingermark right above the lens, so there is no simple answer to your question regarding the point at which the disc should start to spin. Whether or not the deck is on the tilt, the outgoing and returning beams, should not hit any different place on the disc, or the pickup photodiode array. This is the whole purpose of the tracking and focus servos. The lens suspension is very soft, and the mass of the suspension and lens is very small. The focus and tracking motors, which are actually like tiny loudspeakers acting on the lens suspension in all directions, are plenty pwerful enough to hold the lens in the right place, irrespective of gravity acting on the pickup in any direction. When the disc is rotating, you should be able to see the lens moving up and down a little, as it follows the irregularities in the flatness of the disc ( don't believe the specs about disc flatness and eccentricity ... ) You may also be able to see *slight* side to side movement of the lens if the disc is particularly eccentric. To see these movements, you will need to be on the last track of the disc so that the lens is at the disc edge. These are the only movements that you should see with the disc playing normally. The laser is moved on by another motor via a gear reduction system. Approximately once every four to six seconds, you should see the whole laser head move on by about one tooth. This movement should be smooth and regular, and there should not be any overshoot. You certainly should not see any shaking or violent movements of the lens, or pickup, in normal play. If the problem is within the laser, then I think that it would have to be something like a loose critical-angle mirror, that only lays in the right place with the laser tipped, in which case, the only solution would be a replacement laser. What laser type is fitted, as a matter of interest ? Many Sony lasers are really quite cheap. Also, I take it that the disc is clamping correctly to the turntable ? Regarding your question about the three lasers. It does not have three lasers as such. It is a single laser diode, whose output passes through a splitter, to create three beams. The middle one of these is the one which should be centred on the disc's spiral data track. This is accomplished by keeping the two side beams an even distance either side of the track. To do this, these two beams are focused onto two diodes in the pickup array. They are called the "E" diode and "F" diode - hence the servo adjustment "E-F balance". When the middle beam is exactly on the disc track, and the E-F balance adjustment is set correctly, the outputs from the E and F diodes will be equal, and there will be a null servo condition. As soon as the beam moves off track, the E-F output will become unbalanced, and a positive or negative servo condition will be created to correct the beam path. Focus servo operation is likewise based on signal imbalance, but this time as a result of the central beam being focused onto the "A", "B", "C" and "D" diodes. When the beam is correctly in focus, it will be circular, and all four diodes will be equally illuminated. If the beam is out of focus, an astigmatic condition will be created, illuminating two of the diodes by a greater degree, creating an inbalance, which drives the focus servo in the appropriate direction to restore correct focus. Which two diodes are illuminated more, depends on which direction the beam is out of focus. Data output is the result of adding the outputs from these four diodes. Not all optical blocks are 3 beam. Some are single, and use a completely different scheme for deriving the servo information. Arfa |
#18
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
I don't know if I will ever get this working right, but I sure am
learning a lot. I'm sure this has to be looking for a reflection before it decides to starts up the spindle. I can't see the lens with the CD in. If I put a clear protective disc in, nothing happens because it sees no reflection. The unit is a CDP-302, a high end unit. The tag on the laser assy is BU-1 119712 And from its construction, to replace it might cost the same as what I paid for the unit new. Since it will play in any position once started, is it possible the focus/tracking drive isn't working till it starts and the at rest position is causing the reflection to totally miss the target. On Mon, 01 Jan 2007 18:18:05 GMT, "Arfa Daily" wrote: "Sam Goldwasser" wrote in message ... writes: I have run thru the solders on the drive board and elsewhere. The drive is fine when it runs and the "tilt" is a specific angle and repeatable with bumping and jarring having no effect. I am suspecting that it has something to do with the laser itself since the fix (tilt) is so repeatable. It is definitely on the deck. I checked out all the switches and the board on the deck removing and replacing all connectors. Apparently the electronics need feedback from the laser to tell it there is a CD in the deck before it will generate the drive for the spindle. Is this correct??? Therefore the Laser needs to see a reflection. So if the laser is tipped, the reflection won't hit the right spot on the assy. I see only a movement, apparently controlled by coils, that adjusted the laser along the radius of the CD. When it does play, I see the sled shaking back and forth as if following the CD. Shouldn't the laser itself be moving by the coils mentioned above. I think this deck has three lasers to control tracking. I have no idea how this works. I only see one lens. See the CD Player Repair guide at the site below. The sled should not be shaking back and forth, only the lens. The sled should just move occasionally to maintain the lens centered withint its range. I agree with Sam. As to whether the player needs to see a disc on the turntable to spin it up, this depends on the unit. Some players will just go ahead and turn on the laser, and spin up the disc, as soon as a 'laser home' signal is detected. Some coarse speed adjustments then take place, along with focus seek, until the spindle servo starts to see valid data coming off the disc, whereupon, it will lock, and the rotational speed will stabilize to the correct value. Other players will switch on the laser as soon as the home signal is detected, and go ahead and look for a reflection from a disc. Coarse focus search will take place at the same time, until the focus servo locks. At this point, the drive to the spindle motor will be switched on, and the data search will commence. Some players, if they fail to detect an initial reflection, will pulse the spindle motor to rotate the disc some, in case there was a jammy fingermark right above the lens, so there is no simple answer to your question regarding the point at which the disc should start to spin. Whether or not the deck is on the tilt, the outgoing and returning beams, should not hit any different place on the disc, or the pickup photodiode array. This is the whole purpose of the tracking and focus servos. The lens suspension is very soft, and the mass of the suspension and lens is very small. The focus and tracking motors, which are actually like tiny loudspeakers acting on the lens suspension in all directions, are plenty pwerful enough to hold the lens in the right place, irrespective of gravity acting on the pickup in any direction. When the disc is rotating, you should be able to see the lens moving up and down a little, as it follows the irregularities in the flatness of the disc ( don't believe the specs about disc flatness and eccentricity ... ) You may also be able to see *slight* side to side movement of the lens if the disc is particularly eccentric. To see these movements, you will need to be on the last track of the disc so that the lens is at the disc edge. These are the only movements that you should see with the disc playing normally. The laser is moved on by another motor via a gear reduction system. Approximately once every four to six seconds, you should see the whole laser head move on by about one tooth. This movement should be smooth and regular, and there should not be any overshoot. You certainly should not see any shaking or violent movements of the lens, or pickup, in normal play. If the problem is within the laser, then I think that it would have to be something like a loose critical-angle mirror, that only lays in the right place with the laser tipped, in which case, the only solution would be a replacement laser. What laser type is fitted, as a matter of interest ? Many Sony lasers are really quite cheap. Also, I take it that the disc is clamping correctly to the turntable ? Regarding your question about the three lasers. It does not have three lasers as such. It is a single laser diode, whose output passes through a splitter, to create three beams. The middle one of these is the one which should be centred on the disc's spiral data track. This is accomplished by keeping the two side beams an even distance either side of the track. To do this, these two beams are focused onto two diodes in the pickup array. They are called the "E" diode and "F" diode - hence the servo adjustment "E-F balance". When the middle beam is exactly on the disc track, and the E-F balance adjustment is set correctly, the outputs from the E and F diodes will be equal, and there will be a null servo condition. As soon as the beam moves off track, the E-F output will become unbalanced, and a positive or negative servo condition will be created to correct the beam path. Focus servo operation is likewise based on signal imbalance, but this time as a result of the central beam being focused onto the "A", "B", "C" and "D" diodes. When the beam is correctly in focus, it will be circular, and all four diodes will be equally illuminated. If the beam is out of focus, an astigmatic condition will be created, illuminating two of the diodes by a greater degree, creating an inbalance, which drives the focus servo in the appropriate direction to restore correct focus. Which two diodes are illuminated more, depends on which direction the beam is out of focus. Data output is the result of adding the outputs from these four diodes. Not all optical blocks are 3 beam. Some are single, and use a completely different scheme for deriving the servo information. Arfa |
#19
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
Arfa Daily wrote:
"Mark D. Zacharias" wrote in message . net... Leonard Caillouet wrote: "Arfa Daily" wrote in message ... As I said in the previous post, one notable exception to the "don't touch it !" rule, is Pioneer players. The laser power pot is a minuscule little thing located on the laser flexiprint, and is adjusted as part of the setup procedure. Pioneer provide a set of testpoints in a single row on their players, and detailed instructions in their manuals, on how to adjust the entire pot-set, which is often a total of eight pots, and which includes the power pot. Interestingly, Pioneer lasers very seldom fail or wear out. I think in 30 odd years, I have only ever replaced perhaps two of the original types ( there were several, but all pretty similar looking ) and maybe two or three of the later ones which have a couple of tiny pots on their PCB. Arfa This is an interesting point about the Pioneer pickups. Ican't recall ever changing more than one or two myself. I worked at a store that sold tons of Pioneer (Elite, mostly) and since then have not workde on them very much. I can recall cleaning lots of spindle motors, but rarely, if ever changing pickups, Is this the experience of others? What would you attribute the difference in reliability to? Pioneer has always had very good laser products in terms of performance. Do others percieve the reliability to have been very good relative to their competitors as I do? Are the current and lower end products holding up as well? Leonard ----------------------------------------------------------------------- One reason Pioneer pickups were so reliable is that the pickup was mounted "upside down" on many models, thus very little dust etc would settle in on the turning mirror. Still, even their models which played the disc label-up were very reliable so far as pickups were concerned. Except of course for those which had the lens fall off... Mark Z. Hi Mark Have you ever managed to replace a completely missing lens, with one from a different pickup ? I had quite an argument with someone once, who said that it just can't be done, but I do it all the time - I should think at least 4 or 5 a year - did one just 2 weeks ago. I use a lens from a scrap KSS series pickup. I think that the trick to getting it to work - or indeed even replacing the 'real' lens when it's still rolling around in the bottom, is to make sure that //every last trace// of glue is removed from both the lens suspension, and the underside of the lens lip, and then to make certain that the lens is refitted absolutely flat. I then refix it, by putting a couple of tiny drops of cyanoacrylate superglue onto the tip of a tiny flat-end jeweller's screwdy, and letting it seep into the crack by capilliary action. Sometimes, I've found that the focus bias and offset need to be adjusted slightly to optimise the performance back to normal, but mostly, I've found that it just works right off, with no improvement to be gained by adjustment. I'd be interested to know what others' experiences of dealing with this problem are. Arfa I re-attached a lens just the other day, but have never tried using one from another brand pickup. If I don't find the original rolling around in there, I just consider it a lost cause. BTW I have found that Pioneer lasers which oscillate have the lens coming loose but hasn't fallen off yet. I don't use superglue because of the danger of it fogging the lens. Mark Z. |
#20
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() "Mark D. Zacharias" wrote in message t... Arfa Daily wrote: "Mark D. Zacharias" wrote in message . net... Leonard Caillouet wrote: "Arfa Daily" wrote in message ... As I said in the previous post, one notable exception to the "don't touch it !" rule, is Pioneer players. The laser power pot is a minuscule little thing located on the laser flexiprint, and is adjusted as part of the setup procedure. Pioneer provide a set of testpoints in a single row on their players, and detailed instructions in their manuals, on how to adjust the entire pot-set, which is often a total of eight pots, and which includes the power pot. Interestingly, Pioneer lasers very seldom fail or wear out. I think in 30 odd years, I have only ever replaced perhaps two of the original types ( there were several, but all pretty similar looking ) and maybe two or three of the later ones which have a couple of tiny pots on their PCB. Arfa This is an interesting point about the Pioneer pickups. Ican't recall ever changing more than one or two myself. I worked at a store that sold tons of Pioneer (Elite, mostly) and since then have not workde on them very much. I can recall cleaning lots of spindle motors, but rarely, if ever changing pickups, Is this the experience of others? What would you attribute the difference in reliability to? Pioneer has always had very good laser products in terms of performance. Do others percieve the reliability to have been very good relative to their competitors as I do? Are the current and lower end products holding up as well? Leonard ----------------------------------------------------------------------- One reason Pioneer pickups were so reliable is that the pickup was mounted "upside down" on many models, thus very little dust etc would settle in on the turning mirror. Still, even their models which played the disc label-up were very reliable so far as pickups were concerned. Except of course for those which had the lens fall off... Mark Z. Hi Mark Have you ever managed to replace a completely missing lens, with one from a different pickup ? I had quite an argument with someone once, who said that it just can't be done, but I do it all the time - I should think at least 4 or 5 a year - did one just 2 weeks ago. I use a lens from a scrap KSS series pickup. I think that the trick to getting it to work - or indeed even replacing the 'real' lens when it's still rolling around in the bottom, is to make sure that //every last trace// of glue is removed from both the lens suspension, and the underside of the lens lip, and then to make certain that the lens is refitted absolutely flat. I then refix it, by putting a couple of tiny drops of cyanoacrylate superglue onto the tip of a tiny flat-end jeweller's screwdy, and letting it seep into the crack by capilliary action. Sometimes, I've found that the focus bias and offset need to be adjusted slightly to optimise the performance back to normal, but mostly, I've found that it just works right off, with no improvement to be gained by adjustment. I'd be interested to know what others' experiences of dealing with this problem are. Arfa I re-attached a lens just the other day, but have never tried using one from another brand pickup. If I don't find the original rolling around in there, I just consider it a lost cause. BTW I have found that Pioneer lasers which oscillate have the lens coming loose but hasn't fallen off yet. I don't use superglue because of the danger of it fogging the lens. Mark Z. Point taken on the lens fogging. It actually happened to me on one of the first that I tried to reattach some years ago. That is why I now only use the tiniest drop, and let it draw itself into the gap. Fuming and fogging is then not a problem. Next time you get a missing lens, just give a Sony one a try. If it's a lost cause already, you've got nothing further to lose ! Arfa |
#21
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() wrote in message ... I don't know if I will ever get this working right, but I sure am learning a lot. I'm sure this has to be looking for a reflection before it decides to starts up the spindle. I can't see the lens with the CD in. If I put a clear protective disc in, nothing happens because it sees no reflection. The unit is a CDP-302, a high end unit. The tag on the laser assy is BU-1 119712 And from its construction, to replace it might cost the same as what I paid for the unit new. Since it will play in any position once started, is it possible the focus/tracking drive isn't working till it starts and the at rest position is causing the reflection to totally miss the target. I can only imagine that to be the case if the disc was not lying flat on the turntable for some reason. That laser part number is not familiar to me. I'll see if I can find some service info on this player, and see if I can think of anything else that you could reasonably try. It's reaching the point where you have to be starting to think that it's something obscure rather than common. An experienced eye might just look at it, and see immediately what the trouble is. On the other hand, sometimes faults like this can baffle even the most experienced of us. Arfa |
#22
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
writes:
SAM Excellent info! Thanks - This assy appears to have the Laser and photodiode array swapped position wise - IE array is at rt angle to the lens. But that doesn't tell me why I can start the CD in a tilted position then lay it flat and it will play just fine till I stop it and try to restart it. It will only START in the tilted position, but will continue to play in any position. It also must be tilted on power up or insertion to read the TOC, otherwise it won't start on "PLAY" at all. I just can't picture anything inside the laser assy that could be that free to move and also be controlled to lock it into position so that it would continue to play once started. The lens assy only rotates on the plane of the CD and doesnt appear to have movement in the "tip" position. DOES the array need to see anything to cause it to start? Usually needs focus lock to start spinning the disc. DOES it search at all by adjusting the tracking of the lens or the array to FIND the CD? It resets to the inner stop. DOES the array adjust ITSELF to provide tracking? If so how? The entire pickup moves to do tracking. As others have noted, with this dramatic a change, it isn't likely to be a matter of an adjustment, or a bad cable. The servos are no longer capable of maintaining either focus or tracking lock. These are the sort of problems where without a service manual, it's unlikely to be repairable. A new pickup might help, but could also be an electronics problem. --- sam | Sci.Electronics.Repair FAQ: http://www.repairfaq.org/ Repair | Main Table of Contents: http://www.repairfaq.org/REPAIR/ +Lasers | Sam's Laser FAQ: http://www.repairfaq.org/sam/lasersam.htm | Mirror Sites: http://www.repairfaq.org/REPAIR/F_mirror.html Important: Anything sent to the email address in the message header above is ignored unless my full name AND either lasers or electronics is included in the subject line. Or, you can contact me via the Feedback Form in the FAQs. |
#23
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
Sam Goldwasser wrote:
Snip The entire pickup moves to do tracking. The entire pickup moves only when the tracking error reaches a certain point, at which time the pickup is bumped a bit, and the tracking coils re-center. Unless we're talking about an old Philips pivoting pickup with no tracking coils... But I'm pretty sure you knew this anyway. Mark Z. |
#24
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
"Arfa Daily" writes:
wrote in message ... I don't know if I will ever get this working right, but I sure am learning a lot. I'm sure this has to be looking for a reflection before it decides to starts up the spindle. I can't see the lens with the CD in. If I put a clear protective disc in, nothing happens because it sees no reflection. The unit is a CDP-302, a high end unit. The tag on the laser assy is BU-1 119712 And from its construction, to replace it might cost the same as what I paid for the unit new. Since it will play in any position once started, is it possible the focus/tracking drive isn't working till it starts and the at rest position is causing the reflection to totally miss the target. I can only imagine that to be the case if the disc was not lying flat on the turntable for some reason. That laser part number is not familiar to me. I'll see if I can find some service info on this player, and see if I can think of anything else that you could reasonably try. It's reaching the point where you have to be starting to think that it's something obscure rather than common. An experienced eye might just look at it, and see immediately what the trouble is. On the other hand, sometimes faults like this can baffle even the most experienced of us. I agree it may be obscure and possibly an electronic problem. There are some differences in the part of the circuit used for the focus seach, so your hypothesis is at least possible. What I have done in the past to examine what's happening when it tries to focus is to cut down a normal CD to just larger than the inner track. One of those mini-CDs might also work. If you see the lens not moving up and down very far, then initial focus search may be the problme. --- sam | Sci.Electronics.Repair FAQ: http://www.repairfaq.org/ Repair | Main Table of Contents: http://www.repairfaq.org/REPAIR/ +Lasers | Sam's Laser FAQ: http://www.repairfaq.org/sam/lasersam.htm | Mirror Sites: http://www.repairfaq.org/REPAIR/F_mirror.html Important: Anything sent to the email address in the message header above is ignored unless my full name AND either lasers or electronics is included in the subject line. Or, you can contact me via the Feedback Form in the FAQs. |
#25
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
"Mark D. Zacharias" writes:
Sam Goldwasser wrote: Snip The entire pickup moves to do tracking. The entire pickup moves only when the tracking error reaches a certain point, at which time the pickup is bumped a bit, and the tracking coils re-center. Unless we're talking about an old Philips pivoting pickup with no tracking coils... No, not those! Anything but those! ![]() But I'm pretty sure you knew this anyway. Right, it was in direct reponse to the original question but perhaps the reply didn't quite make it intact. :-) --- sam | Sci.Electronics.Repair FAQ: http://www.repairfaq.org/ Repair | Main Table of Contents: http://www.repairfaq.org/REPAIR/ +Lasers | Sam's Laser FAQ: http://www.repairfaq.org/sam/lasersam.htm | Mirror Sites: http://www.repairfaq.org/REPAIR/F_mirror.html Important: Anything sent to the email address in the message header above is ignored unless my full name AND either lasers or electronics is included in the subject line. Or, you can contact me via the Feedback Form in the FAQs. |
#26
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() I re-attached a lens just the other day, but have never tried using one from another brand pickup. If I don't find the original rolling around in there, I just consider it a lost cause. BTW I have found that Pioneer lasers which oscillate have the lens coming loose but hasn't fallen off yet. I don't use superglue because of the danger of it fogging the lens. Mark Z. Point taken on the lens fogging. It actually happened to me on one of the first that I tried to reattach some years ago. That is why I now only use the tiniest drop, and let it draw itself into the gap. Fuming and fogging is then not a problem. Next time you get a missing lens, just give a Sony one a try. If it's a lost cause already, you've got nothing further to lose ! Arfa Hmmm, I've got a PS2 with a lens that has bubbles and scratches in it. I was thinking about trying a lens from a compact stereo CD unit I just shredded for parts. I realize its CD vs DVD but the lens's look the same. As observed "got nothing further to lose !" |
#27
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() tomh ha escrito: Hmmm, I've got a PS2 with a lens that has bubbles and scratches in it. blimey, what do you do to it? set it on fire then put it out by scraping it with a nailfile??!! ;-) |
#28
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
b wrote:
tomh ha escrito: Hmmm, I've got a PS2 with a lens that has bubbles and scratches in it. blimey, what do you do to it? set it on fire then put it out by scraping it with a nailfile??!! ;-) it was "broken" when I got it. The scratches I can see from somebody previously getting in there and fooling around. Which they did. But the bubbles they've got me stumped. If I manage to get a picture of the lens I'll post it. |
#29
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() "tomh" wrote in message news:39Amh.49$y3.6@llnews... b wrote: tomh ha escrito: Hmmm, I've got a PS2 with a lens that has bubbles and scratches in it. blimey, what do you do to it? set it on fire then put it out by scraping it with a nailfile??!! ;-) it was "broken" when I got it. The scratches I can see from somebody previously getting in there and fooling around. Which they did. But the bubbles they've got me stumped. If I manage to get a picture of the lens I'll post it. I've seen bubbled lenses from owners trying cleaning fluids, and also from focus or tracking drive faults, which have heated up the focus or tracking motor coils, resulting in a melted lens. It's worth trying a lens from a different laser ( assuming that there are no drive issues that have melted or seized the lens suspension ). I have never cross-swapped between DVD and CD lenses, but in principle, I don't see why there should be any fundamental difference. For any chance of success though, note carefully what I said about making sure that every last trace of glue is removed from the lens mount, and the replacement lens lip, before fitting it. It is essential that the lens is //absolutely flat// in the mount. Arfa |
#30
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() "Sam Goldwasser" wrote in message ... "Arfa Daily" writes: wrote in message ... I don't know if I will ever get this working right, but I sure am learning a lot. I'm sure this has to be looking for a reflection before it decides to starts up the spindle. I can't see the lens with the CD in. If I put a clear protective disc in, nothing happens because it sees no reflection. The unit is a CDP-302, a high end unit. The tag on the laser assy is BU-1 119712 And from its construction, to replace it might cost the same as what I paid for the unit new. Since it will play in any position once started, is it possible the focus/tracking drive isn't working till it starts and the at rest position is causing the reflection to totally miss the target. I can only imagine that to be the case if the disc was not lying flat on the turntable for some reason. That laser part number is not familiar to me. I'll see if I can find some service info on this player, and see if I can think of anything else that you could reasonably try. It's reaching the point where you have to be starting to think that it's something obscure rather than common. An experienced eye might just look at it, and see immediately what the trouble is. On the other hand, sometimes faults like this can baffle even the most experienced of us. I agree it may be obscure and possibly an electronic problem. There are some differences in the part of the circuit used for the focus seach, so your hypothesis is at least possible. What I have done in the past to examine what's happening when it tries to focus is to cut down a normal CD to just larger than the inner track. One of those mini-CDs might also work. If you see the lens not moving up and down very far, then initial focus search may be the problme. Long ago, I cut a large chunk out of an old CD with a hacksaw. That allowed me to position the hole such that disc was over the lens, but the lens could be seen through the hole. i don't know where that disc is now - buried somewhere deep in the workshop no doubt. These days, I tend to just lift the deck out, and just watch the lens action by looking under the disc or, if you can remove the disc clamp to get good access to the top of the laser, you can just hold a disc over the top of it at about turntable height, and as parallel to the deck as you can manage. You can usually get the focus servo to lock, and even 'play' with it by moving the disc up and down. It's quite fascinating to watch the lens follow the disc as though it was stuck there with invisible glue. I also have a circular magnet removed from a scrap deck's disc clamp. this is very useful for sticking a disc down to the turntable, after removing a disc clamp for eyeball access. You can also BluTack a disc down at a pinch. Having had a further think about this problem, could it be that something is restricting the movement of the sled when the deck is flat ? Most players will shuffle the laser off the home switch and back on again, just to make sure that the laser is home, before turning on the laser and starting the 'disc in' initialisation procedure, ending in reading the TOC. If that shuffle couldn't take place, or was being restricted in some way, then the system control micro may not be getting the right switch signals to start it off. If you can get the deck out in your hand whilst keeping it connected, or otherwise get good visual access to it, try just opening and closing the drawer, with no disc loaded, and watch exactly what happens - laser shuffle, laser burning, focus seek etc, then try the same again with the deck tipped, and see if you can see any difference in what happens. If not, try the same with a disc loaded. Arfa |
#31
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
I've tried so many things, I probably looked at that.
There is no restraint on the sled. It moves freely. The drive belts for the table are weak, but I twisted them and it seemed to work ok. I made sure it was fully seated when closed. It wouldn't clamp the CD if it wasn't. I have watched it seeking track 0 but I don't remember if that was only when tipped or if it did it when it was flat also. I believe there was no action whatever when it is flat, but I will look again. I can't see the lens assy from the bottom, but I will try a small CD and see if I can view it then. thanks. On Wed, 03 Jan 2007 02:19:21 GMT, "Arfa Daily" wrote: "Sam Goldwasser" wrote in message ... "Arfa Daily" writes: wrote in message ... I don't know if I will ever get this working right, but I sure am learning a lot. I'm sure this has to be looking for a reflection before it decides to starts up the spindle. I can't see the lens with the CD in. If I put a clear protective disc in, nothing happens because it sees no reflection. The unit is a CDP-302, a high end unit. The tag on the laser assy is BU-1 119712 And from its construction, to replace it might cost the same as what I paid for the unit new. Since it will play in any position once started, is it possible the focus/tracking drive isn't working till it starts and the at rest position is causing the reflection to totally miss the target. I can only imagine that to be the case if the disc was not lying flat on the turntable for some reason. That laser part number is not familiar to me. I'll see if I can find some service info on this player, and see if I can think of anything else that you could reasonably try. It's reaching the point where you have to be starting to think that it's something obscure rather than common. An experienced eye might just look at it, and see immediately what the trouble is. On the other hand, sometimes faults like this can baffle even the most experienced of us. I agree it may be obscure and possibly an electronic problem. There are some differences in the part of the circuit used for the focus seach, so your hypothesis is at least possible. What I have done in the past to examine what's happening when it tries to focus is to cut down a normal CD to just larger than the inner track. One of those mini-CDs might also work. If you see the lens not moving up and down very far, then initial focus search may be the problme. Long ago, I cut a large chunk out of an old CD with a hacksaw. That allowed me to position the hole such that disc was over the lens, but the lens could be seen through the hole. i don't know where that disc is now - buried somewhere deep in the workshop no doubt. These days, I tend to just lift the deck out, and just watch the lens action by looking under the disc or, if you can remove the disc clamp to get good access to the top of the laser, you can just hold a disc over the top of it at about turntable height, and as parallel to the deck as you can manage. You can usually get the focus servo to lock, and even 'play' with it by moving the disc up and down. It's quite fascinating to watch the lens follow the disc as though it was stuck there with invisible glue. I also have a circular magnet removed from a scrap deck's disc clamp. this is very useful for sticking a disc down to the turntable, after removing a disc clamp for eyeball access. You can also BluTack a disc down at a pinch. Having had a further think about this problem, could it be that something is restricting the movement of the sled when the deck is flat ? Most players will shuffle the laser off the home switch and back on again, just to make sure that the laser is home, before turning on the laser and starting the 'disc in' initialisation procedure, ending in reading the TOC. If that shuffle couldn't take place, or was being restricted in some way, then the system control micro may not be getting the right switch signals to start it off. If you can get the deck out in your hand whilst keeping it connected, or otherwise get good visual access to it, try just opening and closing the drawer, with no disc loaded, and watch exactly what happens - laser shuffle, laser burning, focus seek etc, then try the same again with the deck tipped, and see if you can see any difference in what happens. If not, try the same with a disc loaded. Arfa |
#32
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() wrote in message ... I've tried so many things, I probably looked at that. There is no restraint on the sled. It moves freely. The drive belts for the table are weak, but I twisted them and it seemed to work ok. I made sure it was fully seated when closed. It wouldn't clamp the CD if it wasn't. I have watched it seeking track 0 but I don't remember if that was only when tipped or if it did it when it was flat also. I believe there was no action whatever when it is flat, but I will look again. I can't see the lens assy from the bottom, but I will try a small CD and see if I can view it then. thanks. Well, if the tray drive belt is weak, then the very first thing I would do is to replace it. I wouldn't for one second believe with absolute certainty, that the loading cycle is completing correctly and fully, just because the disc appears to clamp. I have had similar situations many times, where the final push of the mech, beyond just clamping the disc, works a "load complete" switch, and that extra little bit of mechanical resistance from the switch's spring loading, is just enough to hold it open. If, as you suspect that you observed, the deck does absolutely nothing when it is flat, then this would be a good indicator that the system control micro is waiting for something like a "load complete" signal. In this situation, gravity may well take over to assist the last little movement, when the deck is tipped. This would also make sense with your assertion that once the deck has read the TOC 'on the tip' as it were, it will then play in any position. The system control micro only needs to be told that loading is complete, once. Just change the belt, before going any further ... Arfa |
#33
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
Arfa Daily wrote:
"tomh" wrote in message news:39Amh.49$y3.6@llnews... b wrote: tomh ha escrito: Hmmm, I've got a PS2 with a lens that has bubbles and scratches in it. blimey, what do you do to it? set it on fire then put it out by scraping it with a nailfile??!! ;-) it was "broken" when I got it. The scratches I can see from somebody previously getting in there and fooling around. Which they did. But the bubbles they've got me stumped. If I manage to get a picture of the lens I'll post it. I've seen bubbled lenses from owners trying cleaning fluids, and also from focus or tracking drive faults, which have heated up the focus or tracking motor coils, resulting in a melted lens. It's worth trying a lens from a different laser ( assuming that there are no drive issues that have melted or seized the lens suspension ). I have never cross-swapped between DVD and CD lenses, but in principle, I don't see why there should be any fundamental difference. For any chance of success though, note carefully what I said about making sure that every last trace of glue is removed from the lens mount, and the replacement lens lip, before fitting it. It is essential that the lens is //absolutely flat// in the mount. Arfa I don't think a CD lens would EVEN work on a DVD, or vice versa. The inner curvature of the two is different, yes? DVD's have smaller pit size. MarkZ . |
#34
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
NO, I have pushed it to be sure it is all the way in.
The tip is from side to side so gravity is not at work here. A front to back tip will not cause it to operate. Also, it will only CONTINUE to play in any position once play is started. If I tip it to read the TOC then lay it flat it will not start playing. I HAVE discovered something else. The sled rides on 2 rails. one side has a sleeve on it, the other rests on a roller. It will move up about 1/16 inch on that side. If I try to start it while flat AND holding that side up it works. Once started, I have to let it go so the sled can move and it continues to work. Since that is not giving me a 45° tip, it does point to the lens or something inside being too far off to see the reflection in its normal position. Since it comes back at a 45° tip I have to think something inside is loose rather than the lens gimble as that only seems to move in the plane of the CD. HOWEVER, there isn't anything else in the assy that moves except the lens is there?????????? THEREFORE, it has to be the lens gimble, RIGHT????????? On Wed, 03 Jan 2007 10:38:44 GMT, "Arfa Daily" wrote: wrote in message .. . I've tried so many things, I probably looked at that. There is no restraint on the sled. It moves freely. The drive belts for the table are weak, but I twisted them and it seemed to work ok. I made sure it was fully seated when closed. It wouldn't clamp the CD if it wasn't. I have watched it seeking track 0 but I don't remember if that was only when tipped or if it did it when it was flat also. I believe there was no action whatever when it is flat, but I will look again. I can't see the lens assy from the bottom, but I will try a small CD and see if I can view it then. thanks. Well, if the tray drive belt is weak, then the very first thing I would do is to replace it. I wouldn't for one second believe with absolute certainty, that the loading cycle is completing correctly and fully, just because the disc appears to clamp. I have had similar situations many times, where the final push of the mech, beyond just clamping the disc, works a "load complete" switch, and that extra little bit of mechanical resistance from the switch's spring loading, is just enough to hold it open. If, as you suspect that you observed, the deck does absolutely nothing when it is flat, then this would be a good indicator that the system control micro is waiting for something like a "load complete" signal. In this situation, gravity may well take over to assist the last little movement, when the deck is tipped. This would also make sense with your assertion that once the deck has read the TOC 'on the tip' as it were, it will then play in any position. The system control micro only needs to be told that loading is complete, once. Just change the belt, before going any further ... Arfa |
#35
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
Mark D. Zacharias wrote:
Arfa Daily wrote: "tomh" wrote in message news:39Amh.49$y3.6@llnews... b wrote: tomh ha escrito: Hmmm, I've got a PS2 with a lens that has bubbles and scratches in it. blimey, what do you do to it? set it on fire then put it out by scraping it with a nailfile??!! ;-) it was "broken" when I got it. The scratches I can see from somebody previously getting in there and fooling around. Which they did. But the bubbles they've got me stumped. If I manage to get a picture of the lens I'll post it. I've seen bubbled lenses from owners trying cleaning fluids, and also from focus or tracking drive faults, which have heated up the focus or tracking motor coils, resulting in a melted lens. It's worth trying a lens from a different laser ( assuming that there are no drive issues that have melted or seized the lens suspension ). I have never cross-swapped between DVD and CD lenses, but in principle, I don't see why there should be any fundamental difference. For any chance of success though, note carefully what I said about making sure that every last trace of glue is removed from the lens mount, and the replacement lens lip, before fitting it. It is essential that the lens is //absolutely flat// in the mount. Arfa I don't think a CD lens would EVEN work on a DVD, or vice versa. The inner curvature of the two is different, yes? DVD's have smaller pit size. MarkZ . yes point well taken. When I get both lens's out I'll check them carefully. If they do show a remarkable difference I probably have a dead DVD mechanism in my parts/junk pile that I can get a lens from. will advise... tomh |
#36
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() "Mark D. Zacharias" wrote in message t... Arfa Daily wrote: "tomh" wrote in message news:39Amh.49$y3.6@llnews... b wrote: tomh ha escrito: Hmmm, I've got a PS2 with a lens that has bubbles and scratches in it. blimey, what do you do to it? set it on fire then put it out by scraping it with a nailfile??!! ;-) it was "broken" when I got it. The scratches I can see from somebody previously getting in there and fooling around. Which they did. But the bubbles they've got me stumped. If I manage to get a picture of the lens I'll post it. I've seen bubbled lenses from owners trying cleaning fluids, and also from focus or tracking drive faults, which have heated up the focus or tracking motor coils, resulting in a melted lens. It's worth trying a lens from a different laser ( assuming that there are no drive issues that have melted or seized the lens suspension ). I have never cross-swapped between DVD and CD lenses, but in principle, I don't see why there should be any fundamental difference. For any chance of success though, note carefully what I said about making sure that every last trace of glue is removed from the lens mount, and the replacement lens lip, before fitting it. It is essential that the lens is //absolutely flat// in the mount. Arfa I don't think a CD lens would EVEN work on a DVD, or vice versa. The inner curvature of the two is different, yes? DVD's have smaller pit size. MarkZ . I'm not sure about the potential effects of that. Might work one way round, but not the other. A DVD laser, for instance, will happily play a CD or VCD. Arfa |
#37
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() wrote in message ... NO, I have pushed it to be sure it is all the way in. The tip is from side to side so gravity is not at work here. A front to back tip will not cause it to operate. Also, it will only CONTINUE to play in any position once play is started. If I tip it to read the TOC then lay it flat it will not start playing. I HAVE discovered something else. The sled rides on 2 rails. one side has a sleeve on it, the other rests on a roller. It will move up about 1/16 inch on that side. If I try to start it while flat AND holding that side up it works. Once started, I have to let it go so the sled can move and it continues to work. Since that is not giving me a 45° tip, it does point to the lens or something inside being too far off to see the reflection in its normal position. Since it comes back at a 45° tip I have to think something inside is loose rather than the lens gimble as that only seems to move in the plane of the CD. HOWEVER, there isn't anything else in the assy that moves except the lens is there?????????? THEREFORE, it has to be the lens gimble, RIGHT????????? Weeeeeelllllll ... Not necessarily. Firstly, just because you have manually pushed the tray all the way in, that still doesn't say that the whole loading process has completed correctly. There is often continued movement to complete the loading cycle, after the tray is fully home. Sometimes, for instance, a tray lock slides into place, finally working a switch, to tell the micro that loading really is complete. However, leaving that aside for a moment, I think that you can be pretty sure that a 1/16th inch of vertical play at one of the slides, is not correct. Are you absolutely sure that something is not broken or displaced or missing from the mech at that point ? It's not uncommon for there to be a phosphor bronze finger attached to the laser assembly, and resting under the slide rod, to take out just the sort of play that you are describing. If it definitely works under all circumstances with the deck flat, but that side of the laser lifted, then logically, that must be where your problem lies, right ? Arfa |
#38
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
Well, Tipping the deck doesn't move the the sled thru any part of that
1/16 inch, unless it is tipped past 90°. So something else on the sled is moving when it is tipped to cause it to see a reflection. The closed drawer position, I am sure, is not the issue. The sled doesn't move to track zero until it sees a reflection. I can hold it there in a position that otherwise doesn't work (flat or tipped front to back) and it still doesn't start up. The side that is loose appears to be designed that way. There is nothing that would hold it top AND bottom. Gravity keep the top roller in contact with the rail. A piece as you describe would assist gravity to hold it in just this position. There is no adj on that roller to make it bring the sled up. There were no loose pieces in the unit. I might be able to jury-rig this to make the roller bigger and bring the sled up. On Thu, 04 Jan 2007 01:15:18 GMT, "Arfa Daily" wrote: wrote in message .. . NO, I have pushed it to be sure it is all the way in. The tip is from side to side so gravity is not at work here. A front to back tip will not cause it to operate. Also, it will only CONTINUE to play in any position once play is started. If I tip it to read the TOC then lay it flat it will not start playing. I HAVE discovered something else. The sled rides on 2 rails. one side has a sleeve on it, the other rests on a roller. It will move up about 1/16 inch on that side. If I try to start it while flat AND holding that side up it works. Once started, I have to let it go so the sled can move and it continues to work. Since that is not giving me a 45° tip, it does point to the lens or something inside being too far off to see the reflection in its normal position. Since it comes back at a 45° tip I have to think something inside is loose rather than the lens gimble as that only seems to move in the plane of the CD. HOWEVER, there isn't anything else in the assy that moves except the lens is there?????????? THEREFORE, it has to be the lens gimble, RIGHT????????? Weeeeeelllllll ... Not necessarily. Firstly, just because you have manually pushed the tray all the way in, that still doesn't say that the whole loading process has completed correctly. There is often continued movement to complete the loading cycle, after the tray is fully home. Sometimes, for instance, a tray lock slides into place, finally working a switch, to tell the micro that loading really is complete. However, leaving that aside for a moment, I think that you can be pretty sure that a 1/16th inch of vertical play at one of the slides, is not correct. Are you absolutely sure that something is not broken or displaced or missing from the mech at that point ? It's not uncommon for there to be a phosphor bronze finger attached to the laser assembly, and resting under the slide rod, to take out just the sort of play that you are describing. If it definitely works under all circumstances with the deck flat, but that side of the laser lifted, then logically, that must be where your problem lies, right ? Arfa |
#39
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]()
Arfa Daily wrote:
"Mark D. Zacharias" wrote in message t... Arfa Daily wrote: "tomh" wrote in message news:39Amh.49$y3.6@llnews... b wrote: tomh ha escrito: Hmmm, I've got a PS2 with a lens that has bubbles and scratches in it. blimey, what do you do to it? set it on fire then put it out by scraping it with a nailfile??!! ;-) it was "broken" when I got it. The scratches I can see from somebody previously getting in there and fooling around. Which they did. But the bubbles they've got me stumped. If I manage to get a picture of the lens I'll post it. I've seen bubbled lenses from owners trying cleaning fluids, and also from focus or tracking drive faults, which have heated up the focus or tracking motor coils, resulting in a melted lens. It's worth trying a lens from a different laser ( assuming that there are no drive issues that have melted or seized the lens suspension ). I have never cross-swapped between DVD and CD lenses, but in principle, I don't see why there should be any fundamental difference. For any chance of success though, note carefully what I said about making sure that every last trace of glue is removed from the lens mount, and the replacement lens lip, before fitting it. It is essential that the lens is //absolutely flat// in the mount. Arfa I don't think a CD lens would EVEN work on a DVD, or vice versa. The inner curvature of the two is different, yes? DVD's have smaller pit size. MarkZ . I'm not sure about the potential effects of that. Might work one way round, but not the other. A DVD laser, for instance, will happily play a CD or VCD. Arfa That's true. Hmmm. Still, DVD players aren't optimized for CD's... (sound effect of Markie weasling ;-) Mark Z. |
#40
![]()
Posted to sci.electronics.repair
|
|||
|
|||
![]() wrote in message ... Well, Tipping the deck doesn't move the the sled thru any part of that 1/16 inch, unless it is tipped past 90°. So something else on the sled is moving when it is tipped to cause it to see a reflection. The closed drawer position, I am sure, is not the issue. The sled doesn't move to track zero until it sees a reflection. I can hold it there in a position that otherwise doesn't work (flat or tipped front to back) and it still doesn't start up. The side that is loose appears to be designed that way. There is nothing that would hold it top AND bottom. Gravity keep the top roller in contact with the rail. A piece as you describe would assist gravity to hold it in just this position. There is no adj on that roller to make it bring the sled up. There were no loose pieces in the unit. I might be able to jury-rig this to make the roller bigger and bring the sled up. OK. I think that we have now reached the point where we have exhausted all the normal possibilities that any of us would be able to think of with this, which only leaves something very obscure and particular to your player. Having tried just about everything else, I can see nothing else it can be bar an internal defect in the laser. If taking out the play on the slide really does provide a cure, and you're not up for the cost of replacing the laser, which still just *may* not be the cause of the problem, then as it's just for yourself, and not a commercial repair, now might be the time to go for the 'technical bodge' option, as you suggest. If the slide rod os easily removable, how about encasing it in a piece of thin heatshrink tubing ? This would increase the effective diameter of the rod by about the right amount, and give the roller a long life and durable surface surface to run on. Arfa On Thu, 04 Jan 2007 01:15:18 GMT, "Arfa Daily" wrote: wrote in message . .. NO, I have pushed it to be sure it is all the way in. The tip is from side to side so gravity is not at work here. A front to back tip will not cause it to operate. Also, it will only CONTINUE to play in any position once play is started. If I tip it to read the TOC then lay it flat it will not start playing. I HAVE discovered something else. The sled rides on 2 rails. one side has a sleeve on it, the other rests on a roller. It will move up about 1/16 inch on that side. If I try to start it while flat AND holding that side up it works. Once started, I have to let it go so the sled can move and it continues to work. Since that is not giving me a 45° tip, it does point to the lens or something inside being too far off to see the reflection in its normal position. Since it comes back at a 45° tip I have to think something inside is loose rather than the lens gimble as that only seems to move in the plane of the CD. HOWEVER, there isn't anything else in the assy that moves except the lens is there?????????? THEREFORE, it has to be the lens gimble, RIGHT????????? Weeeeeelllllll ... Not necessarily. Firstly, just because you have manually pushed the tray all the way in, that still doesn't say that the whole loading process has completed correctly. There is often continued movement to complete the loading cycle, after the tray is fully home. Sometimes, for instance, a tray lock slides into place, finally working a switch, to tell the micro that loading really is complete. However, leaving that aside for a moment, I think that you can be pretty sure that a 1/16th inch of vertical play at one of the slides, is not correct. Are you absolutely sure that something is not broken or displaced or missing from the mech at that point ? It's not uncommon for there to be a phosphor bronze finger attached to the laser assembly, and resting under the slide rod, to take out just the sort of play that you are describing. If it definitely works under all circumstances with the deck flat, but that side of the laser lifted, then logically, that must be where your problem lies, right ? Arfa |
Reply |
|
Thread Tools | Search this Thread |
Display Modes | |
|
|
![]() |
||||
Thread | Forum | |||
weird toilet problem | Home Repair | |||
Weird A/C problem | Home Repair | |||
Weird TV problem | Electronics Repair | |||
Weird problem | Electronics Repair | |||
weird Refrigerator-TV problem | Home Repair |