I do have to ask, why these meters used BOTH the D cells and other
batteries as well. (In other words, more than one kind of battery). Why
didn't they just use one battery or one set of batteries for the whole
device?


It's for measurement of high resistance values.

Sensitivity of the movement in a 20 kOhms/V AMM is 50 ľA. Current
through a 1.5 MOhm resistor will be 1 ľA with one AA cell (1.5 V):
hard to read on the scale.

With a 15 V battery there will be 10 ľA ...

HTH