"SkyDemon" wrote in message
...
Hi, i have a Kikusui COS5020 oscilloscope and channel 1 on it doesnt work
(not getting any trace at all). Does anyone know where to get the
schematic/service manual for it? (I have found a paid site but ill try and
see if i can get it free first).

Also can anyone point me in the right direction where to look for the

fault?
When no channel is selected there is a faint unfocused mark on the screen
which stays there when ch 1 is selected so this leads me to believe that
maybe there is something wrong with the input selection section.

Hope that made sense

Cheers



Generally speaking the diagnostic problem with
scopes as far as X and Y amps is they have
to be DC coupled throughout. So anywhere in
the chain something failing and sending the
signal high or low will shift the o/p well off beam.
Other than i/p FET overdriven the other common
problem is one of the pair of push-pull output
drivers to the CRT plates having failed.