Default Inappropriate ICD Shocks Caused by External Electrical Noise

Joseph Gwinn writes:

I saw this article in The New England Journal of Medicine. Here is the
URL, unfortunately behind a paywall.


http://content.nejm.org/cgi/content/short/360/13/1363


The gist is that a danish doctor had a patient who was getting
inappropriate cardiac shocks from an ICD (implantable cardioverter-
defibrillator) while in the shower, which turned out to be due to bad
house wiring causing a 50 Hz leakage current from showerhead to floor
drain.


I suspect it was a point-of-use waterheater in the head. Unknown
in the US; they are seen elsewhere.



--