View Single Post
  #2   Report Post  
Posted to sci.electronics.repair
n cook n cook is offline
external usenet poster
 
Posts: 1,220
Default Advice on Gould DSO please...

storan4wd wrote in message
oups.com...
I have just got a Gould 465 DSO (same as 420 & 450 but 100MHz
bandwidth).

With the input grounded the trace is never dead straight (as with an
analogue scope) but exhibits a very, very tiny amount of vertical
noise. It doesn't alter with channel sensitivity and is the same on
both channels, so I'm guessing it's due to ADC error. Everything else
about the scope works perfectly.

It's livable with but if anyone can tell me whether this is normal
behaviour for these scopes (or other DSOs in general) I'd appreciate
it.

Bob


If its invariant noise then surely it must be noise on the final "Y" stage
supply rail. Try adding a HV electrolytic to see if it has any effect.

--
Diverse Devices, Southampton, England
electronic hints and repair briefs , schematics/manuals list on
http://home.graffiti.net/diverse:graffiti.net/