Intensity ripple on Tek 465B
Sam Goldwasser wrote:
So my trusty scope is acting up.
At normal and low intensity, there is a chopped appearance to
the trace at a few hundred Hz. It's variable and at times,
increasing intensity also causes change in focus, and/or the
intensity doesn't go as high as it should.
I assume a problem in the HV.
Thanks for any info.
Maybe check the LV first. The bridges on this scope tend to go
half-bad, turning the bridges into half-wave rectifiers. Also the
filter caps can lose capacitance. Put the scope on a variac, turn the
variac down to 100 volts and put a scope on the PS test points. If you
see 60 Hz ripple, then it's the diodes, if it's 120Hz ripple, its the
capacitors.
The HV power supply has like a voltage quadrupler. Maybe a diode or
two has shorted there?
|