Electronics Repair (sci.electronics.repair) Discussion of repairing electronic equipment. Topics include requests for assistance, where to obtain servicing information and parts, techniques for diagnosis and repair, and annecdotes about success, failures and problems.

Reply
 
LinkBack Thread Tools Search this Thread Display Modes
  #1   Report Post  
Posted to sci.electronics.repair
external usenet poster
 
Posts: 17
Default Tek 7000 Series Fixtures 067-0655-00 vs 067-0589-00 fixtures

We're running through the calibration on our 067-0587-10 Standardizers
and they're failing frequency response. All points lead to the
067-0655-00 fixture going bad, which we only have one of. We tried to
substitute the 067-0589-00 fixture in place of it and pull the + & -
signals off of it, but the frequency response was even worse, on the
order of -12% at points. What is the main difference between the
067-0655-00 and the 067-0589-00? The 0655 looks like they've spent
significant effort shielding the pickoff points, and the 0589 has a
little more unshielded coax showing. However, the flexible extender
board has a similar amount of unshielded coax to the 0589 and seems to
pass the signals just fine. Is it just the rolloff of the coax used
in the 0589?

Are there any engineers out there that worked on the 7000's still
around?

Thanks for any help,

Steve
  #2   Report Post  
Posted to sci.electronics.repair
external usenet poster
 
Posts: 17
Default Tek 7000 Series Fixtures 067-0655-00 vs 067-0589-00 fixtures

For anyone interested, the problem was indeed the jig. There is a "50
Ohm Follower" board that is built into the mainframe, and that is the
primary difference between the 067-0655-00 and the 067-0589-00. We
bypassed the jig by purchasing a used mainframe and pulling the coax
signal lines to the outside of the scope body. It's sad to scrap a
scope like that, but the jig cost ~$3000.00, and a used mainframe is
$300 and has two voltage follower boards. Hope this helps someone,
but I have a feeling there's not a whole lot of 7000 series mainframe
calibration going on anymore.

Steve

On Aug 30, 7:56*am, sck0006 wrote:
We're running through the calibration on our 067-0587-10 Standardizers
and they're failing frequency response. *All points lead to the
067-0655-00 fixture going bad, which we only have one of. *We tried to
substitute the 067-0589-00 fixture in place of it and pull the + & -
signals off of it, but the frequency response was even worse, on the
order of -12% at points. *What is the main difference between the
067-0655-00 and the 067-0589-00? *The 0655 looks like they've spent
significant effort shielding the pickoff points, and the 0589 has a
little more unshielded coax showing. *However, the flexible extender
board has a similar amount of unshielded coax to the 0589 and seems to
pass the signals just fine. *Is it just the rolloff of the coax used
in the 0589?

Are there any engineers out there that worked on the 7000's still
around?

Thanks for any help,

Steve


Reply
Thread Tools Search this Thread
Search this Thread:

Advanced Search
Display Modes

Posting Rules

Smilies are On
[IMG] code is On
HTML code is Off
Trackbacks are On
Pingbacks are On
Refbacks are On


Similar Threads
Thread Thread Starter Forum Replies Last Post
T-12 fixtures again Mike O. Woodworking 19 March 13th 08 05:01 AM
fluorescent fixtures Patt Electronics Repair 6 September 28th 06 03:29 AM
light fixtures zara Home Repair 0 May 5th 06 12:04 AM
Bathroom fixtures PipeDown Home Repair 2 July 8th 05 02:44 PM