Tenma LCR Meter 72-6634 - calibration procedure?
Hi.
I have a second-hand LCR-meter, Tenma model 72-6634, which seems to be reading 30-40% high on the L/C ranges. (R seems accurate, though!) Please can anyone tell me the 'calibration procedure' for this meter? TIA - Mark. |
#2
Posted to sci.electronics.repair
|
|||
|
|||
Tenma LCR Meter 72-6634 - calibration procedure?
MAC wrote in message
... Hi. I have a second-hand LCR-meter, Tenma model 72-6634, which seems to be reading 30-40% high on the L/C ranges. (R seems accurate, though!) Please can anyone tell me the 'calibration procedure' for this meter? TIA - Mark. More likely a repair job, 0.3 to 0.4 percent out for a calibration issue, yours is 2 orders of magnitude over that. The first thing to check is consistency, or not, of the wrong readings for known test items -- Diverse Devices, Southampton, England electronic hints and repair briefs , schematics/manuals list on http://home.graffiti.net/diverse:graffiti.net/ |
#3
Posted to sci.electronics.repair
|
|||
|
|||
Tenma LCR Meter 72-6634 - calibration procedure?
On Sat, 30 May 2009 01:28:16 -0700 (PDT), MAC
wrote: Hi. I have a second-hand LCR-meter, Tenma model 72-6634, which seems to be *seems to be???* So you are not sure? reading 30-40% high on the L/C ranges. (R seems accurate, though!) Please can anyone tell me the 'calibration procedure' for this meter? You can't determine whether it is reading high or not (you do say "Seems to be" but you want to calibrate it? Maybe try taking it to someone who knows how to evaluate it for accuracy (has the requisit standards, for example) then think about having someone calibrate it. TIA - Mark. |
#4
Posted to sci.electronics.repair
|
|||
|
|||
Tenma LCR Meter 72-6634 - calibration procedure?
N_Cook wrote:
MAC wrote in message ... Hi. I have a second-hand LCR-meter, Tenma model 72-6634, which seems to be reading 30-40% high on the L/C ranges. (R seems accurate, though!) Please can anyone tell me the 'calibration procedure' for this meter? TIA - Mark. I've got one of those ! Put in a good battery and recheck it. It eats batteries, at least mine does, and the readings vary greatly with voltage. More likely a repair job, 0.3 to 0.4 percent out for a calibration issue, yours is 2 orders of magnitude over that. The first thing to check is consistency, or not, of the wrong readings for known test items -- Best Regards: Baron. |
Reply |
Thread Tools | Search this Thread |
Display Modes | |
|
|
Similar Threads | ||||
Thread | Forum | |||
Sencore LC53 Manuals, Accessories, and Basic Calibration Procedure - Closing soon! | Electronics Repair | |||
Tenma 72-1040 millivolt meter - need manual | Electronics Repair | |||
Gas Meter Calibration ? | Home Repair | |||
Tektronix TDS540 Calibration Procedure??? | Electronics Repair | |||
Need calibration procedure | Electronics Repair |